Detail NR1 ELEC/INST RM/KL/CM
2024-03-22
L63-RM1-PMR-SERVICE-ANALYZER-
Wahyu Adi Triono
Problem: Flange probe, sample line, filter probe,membran pump dirt of dust coal
Action:
Clean up flange probe, filter probe, sample line, diaphragma pump, disposable filter, dryer gas from dust coal process. Check and clean up filter humidity sensor replace with spare filter if found bad condition ( ADI & EVE RIZKY )

Status: done
Back to report list