Detail NR1 ELEC/INST RM/KL/CM
2025-08-20
463-2K1-&-443-2K1
Wahyu Adi Triono
Problem: PMR Service analyser
Action:
Clean flange probe-probe-sample line-filter sample line-diaphragm pump-filter balston-dryer gas pump from dirty of dust. Check humidity sensor and filter sensor, result function sensor humidity and filter ok ( RIZAL & EVE HADIF )

Status: done