Detail NR1 ELEC/INST RM/KL/CM
2025-10-13
463-2K1-&-443-2K1
Wahyu Adi Triono
Problem: PMR service analyzer
Action:
Clean up flange probe-probe-sample line-filter sample line-diaphragm pump-filter balston-dryer gas pump from dirty of dust. Check humidity sensor function and filter humidity result function humidity sensor normal and filter humidity good condition ( RIZAL &  ANDRI )

Status: done