463-2K1-&-443-2K1
Wahyu Adi TrionoProblem: PMR service analyzer
Action:
Clean up flange probe-probe-sample line-filter sample line-diaphragm pump-filter balston-dryer gas pump from dirty of dust. Check humidity sensor function and filter humidity result function humidity sensor normal and filter humidity good condition ( RIZAL & ANDRI )
Status: done
Back to report list