463-2K1-&-443-2K1
Wahyu Adi TrionoProblem: PMR service analyzer
Action:
Clean up flange probe, probe, filter sample line, sample line, diaphragm pump, filter balston from dirty of dust. Check function humidity sensor and filter humidity, result function humidity sensor good and filter humidity ok ( RIZAL & ANDRI )
Status: done