Detail NR1 ELEC/INST RM/KL/CM
2026-07-13
463-2K1-&-443-2K1
Wahyu Adi Triono
Problem: PMR service analyzer
Action:
Clean up flange probe-probe-sample filter-tube sample line-diaphragm pump-filer balston from dirty of dust use equipment cleaning tools. Check function humidity sensor and filter humidity result function humidity sensor normal and filter good condition ( ADI & ANDRI )

Status: done
Back to report list